All about product
Specification
Reviews
0
Questions
0
Popular
In Stock
Product Code: 1711114995
Brand: Dentsply Sirona
Reviews:
0
Found cheaper?
201 500.00грн.
Buy in 1 click:
Description
Intraoral stationary dental X-ray machine HELIODENT PLUS
Fast and safe X-ray machine for intraoral imaging.
The easy-to-use HELIODENT PLUS will offer you the best image quality combined with the renowned durability of Sirona equipment.
The high-frequency generator minimizes the spectrum of radiation absorbed by soft tissues.
HELIODENT Plus works on direct current, which reduces the radiation load by 30% compared to low-frequency X-ray machines.
Advantages and features
- • Intuitive management
- • The color of the display reflects the different states of the device
- • Strong hinged connection
- • The tube is mounted on the head with a goniometer
Technical characteristics of HELIODENT PLUS
- • Generator type: high frequency DC generator
- • Mains voltage: 240V 10%, 50 – 60 Hz
- • Focal spot size: 0.4 (IEC 336)
- • Voltage on the X-ray tube: 60/70 kV
- • Tube current: 7 mA
- • Exposure time: set from 0.01 to 3.2 seconds
- • Shooting mode: You can freely choose. Switching between two types of sensing devices - film or visiograph
- • Display: automatic activation of the standby mode after the set time.
- • Focus: standard focal distance to the skin is 20 cm. Optional - 30 cm
The colors of the display reflect the different modes of operation
- • Blue - neutral mode, the device is ready for use
- • White - tube cooling
- • Yellow - the emitter is working
- • Red - error message
Dimensions Heliodent plus
Detailed technical characteristics
- Nominal voltage: 200 V - 240 V
- Permissible fluctuations: ± 10%
- Nominal current: 6-5 A
- Nominal frequency: 50 Hz / 60 Hz
- Internal resistance of the network: at 120 V 0.3 Ohm at 200 V - 240 V 0.8 Ohm
- Internal wiring fuse: 16 A inertial
- Input power during irradiation: 1.2 kW
- Input power in standby mode: < 20 W
- Voltage on the tube: 60 kV / 70 kV with the possibility of switching (max. tolerance ± 5 kV)
- Tube current: 7 mA (max. tolerance ± 1.4 mA)
- The shape of the high-voltage curve: high-frequency direct current Residual ripple < 4 kV
- High voltage frequency: 50 kHz - 70 kHz
- Emission time: 0.01 s - 3.2 s (max. tolerance ± 10% +1 ms)
- Ripple/pause ratio: automatic control from 1:1 to 1:60
- General filtering of the X-ray emitter: > 1.5 AI / 70 IEC 60522
- Irradiation field: 0 < 60 mm
- Dose rate: 8.5 mGy/s ±40% at 60 kV 11 mGy/s ±40% at 70 kV
- Measuring instruments: PTW-Nomex with an ionization space of 1 cm3 or Unfors multimeter
- Measurement conditions: 200 mm focal length of the device 230 V Nominal voltage
- Focal spot size according to IEC 60336: 0.4
Specifications
Characteristics
Reviews
There are no reviews for this product, be the first to leave your review.
Answers & questions